The Corentium radon gas monitor is a state-of-the-art measuring instrument that combines ease of use and performance.

The Corentium monitor allows you to take a reading of the radon levels and its LCD screen displays the average daily, weekly and long term concentrations. Powered by 3 standard AAA batteries, the monitor makes it easy to take measurements from one room to another in order to get an overview of the concentrations of radon in a home, workplace, school, a daycare centre or any other location.  The Corentium monitor is also an essential tool to use when performing property inspections or when checking the operation of a radon mitigation system.

The internal architecture of the Corentium monitor comes from advanced technology normally reserved for commercial-type detectors. Particular attention has been paid to the quality and protection of the internal components, allowing the use of algorithms for signal analysis that are more sophisticated. The Corentium monitor quickly adapts to its environment and eliminates inaccuracies related to external factors, allowing it to obtain the best accuracy in its niche.

The Corentium monitor is guaranteed for one year and in normal home use, it requires no annual calibrations throughout its useful life, estimated at over 10 years.

 

Corentium2

Specification

 

Radon Sampling Method Passive radon diffusion chamber
Radon Detection Method Alpha spectrometry using digital detector technology.
Precision (at 100Bq/m³)
Short term (7 days)
Long term
20% after 1 week
10% after 1 month
Accuracy < 5% ± 5 Bq/m3
Power Supply 3 AAA alkaline battery
3 years battery life-time
Power Consumption < 250µW
Dimensions 120mm × 69mm × 22.5mm
Weight 130 grams (incl. batteries)
Operation Environment Temperature: 0 °C to +40 °C
Relative Humidity < 95%
Measurement range 0 Bq/m³ (lower detection limit)
9999 Bq/m³  (upper display limit)

Certification

Radon monitors should preferably undergo international comparative tests, in order to compare against other technologies and producers. Since there are so few similar digital monitors in the marked, the Corentium monitor has so far only been able to be tested against traditional alpha track detectors.

The Corentium monitor (named ‘Canary’ in the European market) was in July 2011 tested at the National Institute of Radiological Sciences (NIRS) in Japan. NIRS has calibrated laboratories, and are certified to conduct such tests. The methodology used is that we send a number of monitors to the laboratory; they conduct a test at (for us) an unknown radon level (blind test), and return the monitors to us. We then report our results to them, before they present the correct values, i.e. their measured results. For alpha track detectors, typically 70% of the participants have a result within +-20% of the reference value. This implies that measuring with an alpha track detector in a room with exactly 100 Bq/m3 over a two months period, the measurements done in that room with an alpha track detector from a random supplier will show values from 60 Bq/m3 up to 140 Bq/m3 with 95% probabilityIn the test at NIRS, Canary had a deviation of 3% of the NIRS reference value. Since the reference value has an uncertainty of 6%, the Canary is thus well within the NIRS laboratory measurement uncertainty.

This means that for a 2 month measurement period at 100 Bq/m3 – as discussed above regarding the alpha track detector – one can expect that a random Corentium monitor shows between 90Bq/m3 and 110Bq/m3.

The NIRS report may be downloaded here: NIRS report

In September 2012 Canary was tested in the Federal Office for Radiation Protection (Bfs) in Germany. 21 monitors were tested against reference monitors, and all were statistically measured to be within the laboratories own measurement uncertainty – which is 7%.

The BfS report may be downloaded here: BfS report

In June 2013 the Corentium was again tested at BfS. 6 monitors were exposed for 1100 h*kBq/m3 and showed an average of 3.6% below the BfS reference value.

The BfS report may be downloaded here: BfS report

Corentium was in 2013 tested at IRSN (Institut de Radioprotection et de Sûreté Nucléaire) in France.  The test was conducted over 3 months at a radon level of 170 Bq/m3. The average for the 20 tested devices was 167 Bq/m3.

The IRSN report may be downloaded here: IRSN report

Corentium has also been tested in calibrated radon laboratories in the Czech Republic. These tests verify that Corentium shows the same radon values irrespective of any changes in temperature, humidity, aerosols (dust particles in the air) and electromagnetic fields. This is in contrast to many monitors that are sensitive, and will typically show wrong results when these parameters change.

Corentium5

Technology

Radon monitoring now is taking a permanent step into the digital age – also for the home owner. It has become much easier to diagnose the health of buildings when it comes to radon radiation. Radon measurement procedures have become much easier, more flexible, and more accurate.

The use of Corentium does not require any knowledge about the monitor technology. For those eager to have more insight into the world of digital radon measurement, we will go into more detail here:

Measurement principles

A Corentium can be compared to a digital version of a traditional alpha track detector. The detection of radon is based on the principle that radon gas diffuses into a detection chamber. When the radon atoms decay they emit energetic alpha particles. The alpha particles are detected by a silicon photo diode. Every alpha particle generates a small signal current when it hits the photo diode. By the use of a low-power amplifier stage the signal current is converted into a larger voltage signal. The maximum amplitude of the voltage signal is detected and sampled by an analog to digital converter (ADC). The amplitude is proportional to the energy of the alpha particle that hit the photo diode. The brain of the monitor is a micro-controller that registers the time and energy of every detected particle. This information is used to calculate the mean radon gas concentration for daily, weekly and yearly periods. This is all realized in electronic circuitry that in total only consumes a few microwatts of power, making the Corentium able to operate for almost 3 years on a single set of small AAA alkaline batteries.

The calculation of radon gas concentration is complex since there are multiple error sources that have to be considered. When radon decays, several radioactive so-called daughters are produced – most notably Polonium-218 and Polonium-214, – which themselves emit alpha particles like radon does. These alpha particles could be mistaken for alpha particles from radon and give a false radon gas reading, but Corentium uses algorithms that are able to distinguish between the various alpha sources. The main input to these algorithms is the alpha particle energy in combination with the knowledge of the energy spectrum of radon and polonium isotopes. The Corentium has a very good energy measurement resolution, which is required when using such algorithms.

Accuracy and precision

The Corentium is not constructed to give instantaneous measurement results of the radon gas concentration, but rather averages over time. The reason is that instantaneous values are of little importance, since radon can vary a lot over a short time span, and since the health effect is linked to the average exposure over time. This is the reason why the three values that the Corentium presents are referred to as ‘averages’. The only difference in these three averages is the time span over which the averaging takes place. The day average (1 day) is taken over the last 24 hours, the week average (7 day) over the last 7 days and the long term average for the duration since the Corentium was first started (or last RESET). If the total measurement period is more than one year, the long term measurement is the average over the last year only.

The Corentium performs several levels of calibration. The first calibration takes place during the first 3 minutes after the batteries are inserted. After the Corentium has been running for several days it will ‘learn’ the local conditions and measurement accuracy will improve over time. As an example, the daily averages presented will be more precise after some weeks than during the first few days. The reason is that measurements are affected by aerosols and trace gases in the air, which are typically ‘stable’ for each house/environment and changes little over time, allowing the Corentium to ‘learn’ the local conditions. When the Corentium is first started there is no history for the current location, and the display shows a value of 0 Bq/m3 (15 Bq/m3 in the first sold version of the Corentium). As an example, if the unit is placed in a room with exactly 100 Bq/m3, one will see the measurement slowly ramp up from 0 (15 Bq/m3 in the first sold version of the Corentium) to 100 Bq/m3 during the first 24 hours. As mentioned earlier, the unit does not measure an instantaneous radon gas concentration, but rather an average over a time interval. Therefore, it is of little value to read the unit in the first few hours after it was first started – one should give it at least 24 hours. Before that, the values are likely to be an underestimation of the real value. However, in the case where a room has a very high radon gas concentration – say 1000 Bq/m3 as an example – even the underestimation you get after a few hours will give a hint that the unit may reach high values by the end of the first day.

There is a similar situation related to the weekly average. Until the Corentium has been measuring a full week, the weekly average value will not be as precise as when the unit has been measuring for more than a week.

At this point it is natural to bring forward the two concepts; measurement accuracy and measurement precision.

Accuracy is related to how well the Corentium measurement value matches the true radon gas concentration. This is verified by placing many Corentium units in a laboratory with a known radon gas concentration to check how well the Corentium results match those of professional high cost reference monitors (which is the proxy for the ‘true’ radon gas concentration in the laboratory). We have performed such measurements in renowned laboratories in Japan and the Czech Republic. The result of these tests shows that the Corentium accuracy is around 5%. This means that the results of such comparison measurements are within +/-5% of the result the reference monitors show. Accuracy is really related to systematic errors, for example due to a wrong calibration or other effects that systematically would produce too high or too low values.

Precision on the other hand is related to the spread of the measurement results. If we would place 100 Corentium units side by side in the same room (i.e. at the same radon gas concentration), how well do the results of the individual Corentium units match each other? Looking back at accuracy, this would be how well the average readings of these 100 units match the true radon gas concentration (or how it matches a professional well calibrated monitor). Since we have indicated that Corentium has 5% accuracy, it means we can assume that the average of the 100 monitors would be no more than 5% away from the true value.

For the Corentium the measurement precision (spread in measurement values for the various Corentium) depends on the radon level itself. The higher the radon gas concentration, the more precise the measurement will be. The precision is also different for the various time averages that are measured. The longer the measurement duration the more precise the result will be. These measurement precisions are illustrated in the tables below.

It is customary to give the measurement results (for radon) with a 20% precision. When a precision is indicated it will for most monitors mean that the measurement will follow a Gaussian distribution. A Gaussian distribution with a 20% standard deviation means that in about 68% of the cases (say you had 100 units) the result of a selected unit (pick one of the 100) will be inside a window of +/-20% of the average in 68% of the cases. If the average measurement of the 100 units were 100 Bq/m3, it means you can expect 68 of them to show a result from 80 to 120 Bq/m3. Further, the Gaussian distribution means that approximately 95% of the measurements will be within two times the standard deviation (i.e. 40%) of the average. This means for a 20% reported precision you can expect 95 of the 100 units to show a value between 60 and 140 Bq/m3 for a true radon concentration in the room of 100 Bq/m3.

The total measurement uncertainty is typically the root mean square of the individual contributions (where individual contributions are Gaussian distributed). If we operate with an accuracy of 5% and a precision of 20%, the fact that the accuracy is much better than the precision will only slightly increase the total uncertainty above the precision:

20%2+5%221%

This means we are dominated by the uncertainty introduced by the lower precision.

In a large comparison study between various suppliers of alpha track detectors performed by NIRS (The Japanese Health Authorities) in 2009 it was found that 16 of the 26 contributors (62%) were within +/-20% of the reference value reported by NIRS. Comparing this result to the earlier discussions about precision, we see that this can be taken to indicate that alpha track detectors typically have a measurement precision of about 20%. This is the reason why our tables below show typical measurement durations and radon gas concentration which will allow the Corentium measurement precision to be 20% or better (than alpha track detectors).

The following tables show the total measurement of uncertainty for a Corentium for different radon gas concentrations and for the various averages (weekly and long term). The table does not present values for radon gas concentrations below 100 Bq/m3, since no countries use an action level below this value.

Corentium

Measurement duration to get below the indicated uncertainty for the long term measurement

Radon concentration

[Bq/m3]

20% uncertainty 10% uncertainty
100 4 days 3 weeks
200 2 days 10 days
300 2 days 7 days
500 1 day 4 days
1000 1 day 2 days
2000 1 day 1 day

Measurement uncertainty (%) for the weekly average at different radon gas levels

Radon gas concentration

[Bq/m3]

Measurement uncertainty

[%] 

100 16
200 12
300 10
500 8
1000 7
2000 6

 

Lifetime

The monitor is tested and quality assured at production. It meets the accuracy specified in the specification table, unless it is continuously measured at high radon levels (many thousand Bq/m3) over years. We recommend that the unit is left on all the times.

Corentium3 Corentium4